- Monitor Fridge / Freezer Temps Simultaneously. Email / SMS Alerts. Web Dashboard No Cloud. SoftwareLess PDF Reports.
Data Loggers, Vaccine Temperature Monitors
- Logs: 24M
- Range: -40/50°C
- Accuracy: ± 0.5°C
- Model: ACCSL2021
Home » Products » Vaccine Temperature Monitors » UTREL16 Ultra-Low Freezer Temp Data Logger
The UTREL30-16 30 day logger with display.
A CDC / VFC compliant ISO/IEC 17025 NIST traceable temperature data logger for monitoring & recording temperatures of vaccines in refrigerators or freezers.
New large Easy-to-Read display to see the current, min and max temperatures and alarm status for up to 30 days at a glance.
The UTREL30-16 measures and stores up to 16,129 readings without needing off loading of data — that’s 5 months of recordings @ 15 minute intervals.
Check and reset min-max temperature readings at any time.
Audible and Flashing LED signals temperature excursions. Download data via LogTag Analyzer Software or a PDF (no software required).
UTREL30-16 has a wide temperature range of -90°C to +40°C (-130°F to +104°F).
Software
LogTag Analyzer User Guide… Software Revision: 3.0 Document Revision: 2 Published: 12 January 2018
LogTag Analyzer 3.2.2
LogTag download page for additional software and drivers
CATALOG NO. | DESCRIPTION |
---|---|
LTUTREL3016 | Vaccine Temperature Data Logger – 16K logs w/Stainless Steel Air Probe |
FEATURES:
SPECIFICATIONS:
Product Model | LTUTREL3016 |
External Sensor Measurement Range | -90°C to +40°C (-130°F to +104°F). |
Operating Temperature Range | -30°C to +60°C (-22°F to +140°F). |
Storage Temperature Range | 0°C to +40°C (32°F to +104°F). |
Rated Temperature Reading Resolution | 0.1°C (0.2°F) for measurements -80°C to 10°C (-112°F to 50°F) 0.2°C (0.4°F) for measurements 10°C to 20°C (50°F to 68°F) and -90°C to -80°C (-130°F to -112°F) 0.3°C (0.5°F) for measurements 20°C to 30°C (68°F to 86°F) 0.5°C (0.9°F) for measurements 30°C to 40°C (86°F to 104°F) LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®. |
Sensor Reaction Time | Typically less than 2 minutes(T90) in moving air (1m/s) for ST10S type |
Rated Temperature Reading Accuracy | Better than ±0.5°C for -20°C to +10°C (-4°F to 50°F) Better than ±0.7°C for -50°C to -20°C (-58°F to -4°F) and +10°C to +30°C (50°F to 86°F) Better than ±1°C for -80°C to -50°C (-112°F to -58°F ) and +30°C to +40°C (86°F to 104°F) Better than ±1.8°C for -90°C to -80°C (-130°F to -112°F) |
Recording Capacity | 16,129 temperature readings. 67 days at 6 minute logging. Day summary statistics memory (for display on LCD): up to 30 days of Min/Max/duration and alarm trigger statistics. |
Sampling Interval | Conigurable from 30 seconds to 18 hours. |
Logging Start Options | Push button start (with conigurable start delay from 1 minute to 72 hours) or speciic date & time. |
Download Time | Typically with full memory (16,129 readings) in less than 10 seconds. |
Environmental | IP51 (when vertically mounted or hung with USB dust-cap in place). |
Power Source | 5V via micro USB socket or 2 x 1.5V AAA batteries. CR2032 LiMnO2 coin cell (backup logging function only). |
Battery Life | AAA battery up to 12 months – if unit is operated without power supply. |
Real Time Clock | Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C). |
Probe Cable Length | 3m (9’10”) (recommended maximum). |
Remote Sensor Cable Type | PTFE (FDA food contact rated) coaxial. |
Connection Interface | USB 2.0 / Micro USB Socket. |
Size | 123mm(H) x 63mm(W) x 12.5mm(T). |
Weight | 250g with AAA batteries, (80g – excluding probe and AAA batteries). |
Case Material | Polycarbonate. |
Not applicable
CALL: (973) 300-9100
9am to 5pm est
The UTREL30-16 30 day logger with display.
A CDC / VFC compliant ISO/IEC 17025 NIST traceable temperature data logger for monitoring & recording temperatures of vaccines in refrigerators or freezers.
New large Easy-to-Read display to see the current, min and max temperatures and alarm status for up to 30 days at a glance.
The UTREL30-16 measures and stores up to 16,129 readings without needing off loading of data — that’s 5 months of recordings @ 15 minute intervals.
Check and reset min-max temperature readings at any time.
Audible and Flashing LED signals temperature excursions. Download data via LogTag Analyzer Software or a PDF (no software required).
UTREL30-16 has a wide temperature range of -90°C to +40°C (-130°F to +104°F).
Software
LogTag Analyzer User Guide… Software Revision: 3.0 Document Revision: 2 Published: 12 January 2018
LogTag Analyzer 3.2.2
LogTag download page for additional software and drivers
CATALOG NO. | DESCRIPTION |
---|---|
LTUTREL3016 | Vaccine Temperature Data Logger – 16K logs w/Stainless Steel Air Probe |
FEATURES:
SPECIFICATIONS:
Product Model | LTUTREL3016 |
External Sensor Measurement Range | -90°C to +40°C (-130°F to +104°F). |
Operating Temperature Range | -30°C to +60°C (-22°F to +140°F). |
Storage Temperature Range | 0°C to +40°C (32°F to +104°F). |
Rated Temperature Reading Resolution | 0.1°C (0.2°F) for measurements -80°C to 10°C (-112°F to 50°F) 0.2°C (0.4°F) for measurements 10°C to 20°C (50°F to 68°F) and -90°C to -80°C (-130°F to -112°F) 0.3°C (0.5°F) for measurements 20°C to 30°C (68°F to 86°F) 0.5°C (0.9°F) for measurements 30°C to 40°C (86°F to 104°F) LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®. |
Sensor Reaction Time | Typically less than 2 minutes(T90) in moving air (1m/s) for ST10S type |
Rated Temperature Reading Accuracy | Better than ±0.5°C for -20°C to +10°C (-4°F to 50°F) Better than ±0.7°C for -50°C to -20°C (-58°F to -4°F) and +10°C to +30°C (50°F to 86°F) Better than ±1°C for -80°C to -50°C (-112°F to -58°F ) and +30°C to +40°C (86°F to 104°F) Better than ±1.8°C for -90°C to -80°C (-130°F to -112°F) |
Recording Capacity | 16,129 temperature readings. 67 days at 6 minute logging. Day summary statistics memory (for display on LCD): up to 30 days of Min/Max/duration and alarm trigger statistics. |
Sampling Interval | Conigurable from 30 seconds to 18 hours. |
Logging Start Options | Push button start (with conigurable start delay from 1 minute to 72 hours) or speciic date & time. |
Download Time | Typically with full memory (16,129 readings) in less than 10 seconds. |
Environmental | IP51 (when vertically mounted or hung with USB dust-cap in place). |
Power Source | 5V via micro USB socket or 2 x 1.5V AAA batteries. CR2032 LiMnO2 coin cell (backup logging function only). |
Battery Life | AAA battery up to 12 months – if unit is operated without power supply. |
Real Time Clock | Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C). |
Probe Cable Length | 3m (9’10”) (recommended maximum). |
Remote Sensor Cable Type | PTFE (FDA food contact rated) coaxial. |
Connection Interface | USB 2.0 / Micro USB Socket. |
Size | 123mm(H) x 63mm(W) x 12.5mm(T). |
Weight | 250g with AAA batteries, (80g – excluding probe and AAA batteries). |
Case Material | Polycarbonate. |
Not applicable